Chapter 5: Using The Procedure Station
Product Service 1-800-522-7658
36
Original instructions
Working In the Procedure Station
Note: A more thorough review of using a Procedure station can be found in:
Biosafety in Microbiological and Biomedical Laboratories
(BMBL), Published by the Centers for Disease Control and Prevention
(www.cdc.gov/biosafety/publications
).
Planning
Thoroughly understand procedures and equipment required before
beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the
room while the cabinet is in use.
Start-up
Turn off UV light if included.
Slowly raise the sash until the bottom of the sash aligns with the sash
indicator decal located on the left side of the work area.
Turn on the fluorescent light and cabinet blower if the Smart-Start
features have not been activated.
Check the air grilles for obstructions.
Allow the cabinet to operate until the display screen is shown.
Warming up Screen Display Screen
Wash hands and arms thoroughly with germicidal soap.
Wear appropriate personnel protective equipment (PPE).
Wipe-Down
Raise the sash to its full open position (approximately 21.75 inches or
552 mm). Mute the alarm by depressing the "OK/Mute" switch.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.