ADJUSTING THE BLOWER SPEED, FILTER GAUGE,
AIRFLOW SENSOR AND ALARM SETPOINT
Type B2
* This screen unavailable on units with firmware prior to 10443 Rev. G.
Press the or keys to adjust the blower
speed. The wordwait will flash until the
b lo we r sp e e d st a b ilize s. Pre ss ‘ O K’ to
accept the new setting.
CERTIFIER USE ONLY
Calibration mode:
Motor Speed Factor:
wait 900
950
UP/ DO WN TO CHANGE, OK
TO NEXT, MENU TO LAST
80
C ERTIFIER USE O NLY
Calibration mode:
Filter setup:
1. New Filter
2. Filter Not Changed
UP/DO WN TO CHANGE, OK
TO NEXT, MENU TO LAST
Press the or keys to select. Option # 1 will
reset the filter life remaining gauge to 100%.
Option # 2 will leave the gauge unchanged.
Press ‘OK’ to ac cept your choic e.
CERTIFIER USE ONLY
Calibration mode:
Inf low:
FPM
UV to autoset
UP/ D O WN TO CHANGE, OK
TO NEXT, MENU TO LAST
105
Press the or keys to adjust the
inflow to match the c ertified value.
Press ‘UV Light’ to reset the inflow to
105 FPM. Press ‘OK to accept.
Close the exhaust damper until exhaust
volum e equals 80% of nom inal volume.
PressOK to accept the alarm volume.
Reopen the damper to its normal position.
CERTIFIER USE ONLY
Calibration mode:
reduce inflow to
alarm level,
OK TO SET
MENU TO SKIP
*
The Logic will automatically
turn off and reset itself into
normal operation.