ADJUSTING THE BLOWER SPEED, FILTER GAUGE
AND AIRFLOW SENSOR
Type A2 with airflow sensor
Press the or keys to adjust the blower
speed. The word ‘wait’ will flash until the
blower speed stabilizes. PressOK to
accept the new setting.
CERTIFIER USE ONLY
Ca lib rat io n m od e:
Motor Speed Fac tor:
wait 900
950
UP/ D O WN TO CHANGE, OK
TO NEXT, MENU TO LAST
80
CERTIFIER USE ONLY
Calibration mode:
Filter setup:
1. New Filter
2. Filter Not Changed
UP/ DO WN TO CHANGE, OK
TO NEXT, MENU TO LAST
Press the or keys to select. Option # 1 will
reset the filter life remaining gauge to 100%.
Option # 2 will leave the gauge unchanged.
PressOK to accept your choice.
The Logic will automatically
turn off and reset itself into
normal operation.
C ERTIFI ER USE O N LY
Calibration mode:
Downflow:
FPM
UV to autoset
UP/ D OWN TO CHANGE, OK
TO NEXT, MENU TO LAST
53
CERTI FI ER USE O N LY
Calibration mode:
Inflo w:
FPM
UV to autoset
UP/ D O WN TO CHANGE, OK
TO NEXT, MENU TO LAST
105
Press the or keys to adjust the
inflow to match the c ertified value.
PressUV Light to reset the inflow to
105 FPM. PressOK to accept.
Press the or keys to adjust the
downflow to match the certified value.
Press ‘UV Light to reset the inflow to
55 FPM. Press ‘OK’ to accept.