Quickstart
How do I set up automatic operation of the blower and
lights?
1. Press the
Menu
button.
2. Press the button until the
MyLogic
option is highlighted it will
turn blue).
3. Press
OK/Mute
to enter the MyLogic menu screen.
4. Press the button until the option
configure my Logic for use
option is highlighted (it will turn white).
5. Press the button to select either Smart-Start or manual
operation of the cabinet blower. (In Smart-Start, the cabinet
blower will automatically start every time the sash is opened.
6. Press the button to select either Smart-Start or manual
operation of the fluorescent lights.
7. Press the button to select either Night-Smart or manual
operation of the cabinet blower. (In Night-Smart, the cabinet
blower will automatically operate at idle every time the sash is
closed.
8. Press the button to select either Night-Smart or manual
operation of the UV Light.*
9. Press the button to select the time interval the UV light will
operate for until it automatically shuts off.*
* - These screens will only appear if the unit is configured for a UV light.