Appendix D: References
ASHRAE 110-1995 Method of Testing Performance of
Fume Hoods
Evaluates fume hood’s containment characteristics
Three part test: Smoke generation, Face velocity profile,
Tracer gas release @ 4 liters per minute
Rated As Manufactured (AM), As Installed (AI) and As
Used (AU)
American Society of Heating, Refrigerating, and Air
Conditioning Engineers
1791 Tullie Circle N.E.
Atlanta, GA 30329
(404) 636-8400
ANSI Z9.5-1993 Laboratory Standard
Covers entire laboratory ventilation system.
Vertical stack discharge @ 2000-3000 fpm
New and remodeled hoods shall have a monitoring
device
Ductless hoods should only be used with non-hazardous
materials
Fume hood face velocities between 80-120 fpm
American Industrial Hygiene Association
2700 Prosperity Avenue, Suite 250
Fairfax, VA 22031
(703) 849-8888
SEFA 1-2002
Fume hood face velocities based on toxicity levels of
chemicals
Class A – 125 to 150 fpm
Class B – 80 to100 fpm
Class C – 75-to 80 fpm
Test method – face velocity profile and smoke
generation
Scientific Equipment & Furniture Association
1028 Duchess Drive
McLean, VA 22102
(703) 538-6007
NFPA 45 – 2002 Fire Protection for Laboratories Using
Chemicals
Laboratory hoods should not be relied on for explosion
protection
Exhaust air from fume hoods should not be recirculated
Services should be external to the hood
Canopy hoods only for non-hazardous applications