User guide
9
Product Service 1-800-522-7658
Chapter 2: Prerequisites
W/ Rear Lower Vertical Sash Closed and Front
Vertical Sash Open or Horizontal Doors Open
Hood Size Standard Model Description
Face
Velocity
Exhaust
Volume
Static
Pressure
Loss
Sashes
Open
4' Vertical 48" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 550 CFM 0.10" H
2
O
100 fpm 680 CFM 0.16" H
2
O
1
5' Vertical 60" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 720 CFM 0.16" H
2
O
100 fpm 900 CFM 0.24" H
2
O
1
6' Vertical 72" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 900 CFM 0.18" H
2
O
100 fpm 1120 CFM 0.28" H
2
O
1
8' Vertical 96" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 1250 CFM 0.12" H
2
O
with Vertical-Rising Sashes 100 fpm 1560 CFM 0.19" H
2
O
1
8' Horizontal 96" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 1760 CFM 0.19" H
2
O
with horizontal-sliding sashes 100 fpm 2200 CFM 0.30" H
2
O
2
10' Horizontal 120" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 1090 CFM 0.08" H
2
O
with horizontal-sliding sashes 100 fpm 1360 CFM 0.12" H
2
O
1
12' Horizontal 144" Protector XL Floor-Mounted (Walk-In) Hood 80 fpm 1330 CFM 0.10" H
2
O
with horizontal-sliding sashes 100 fpm 1660 CFM 0.15" H
2
O
1
16' Horizontal 192" Protector XL Floor-Mounted (Walk-In) Hood
with horizontal-sliding sashes 100 fpm 2260 CFM 0.10" H
2
O
1
NOTE: 8' Floor-Mounted Hoods are offered with both
vertical-rising and horizontal-sliding sashes, and have
different CFM’s and pressures.
Proper blower selection can be determined from these
exhaust requirements and the total system static
pressure loss. Contact Labconco Customer Service at
800-821-5525 for assistance in sizing a blower system.