Manual
Chapter 5: Using Your Glove Box
Product Service 1-800-522-7658
31
Calibration Program Screens
Enter code for
certification use only and
then press
ENTER key
CALIBRATION LOCK
?? to Choose, ENTER
to Select, MENU to Exit
0
A
IRFLOW SENSOR ZERO
Please Wait:
150
A
IRFLOW
SENSOR LEVEL ADJUST
Set to Known
A
irflow
?? to Choose, ENTER to Select,
MENU to Exit
A
irflow:
70
CFM
CALIBRATION
?? to Choose,
ENTER to Select,
MENU to Exit
f
1.
A
irflow Senso
r
2.
A
irflow Alarm
3. Pressure Senso
r
4. Pressure Alarm
5. Box Size
6. Box Type
A
LARM LEVEL
?? to Choose, ENTER
to Select, MENU to Exit
A
larm Level: 10
CFM
A
larm Level is best set at low levels such
as 10 CFM below oper
ating speed.
PRESSURE SENSOR
1. Zero Adjust
2. Level Adjust
PRESSURE ALARM
?? to Choose, ENTER to
Select, MENU to Exit
A
larm Level: X.
XX
IOW
PRESSURE SENSOR
ZERO ADJUST
Open all doors
Enter when done
PRESSURE SENSOR
ZERO ADJUST
Wait
150
Seconds
PRESSURE SENSOR
LEVEL ADJUST
Set known Pressure
?? to Choose, ENTER to
Select, MENU to Exit
Pressure: X.XX
IOW
A
IRFLOW ALARM
1
.
A
larm Level
2. Alarm Delay
BOX SIZE
1
. Single
2. Double
A
IRFLOW SENSOR
1
. Zero Adjust
2. Level Adjust
A
LARM DELAY
?? to Choose, ENTER to
Select, MENU to Exit
A
larm Level: x
xx
Seconds
PRESSURE ALARM
1
.
A
larm Level
2. Alarm Delay
A
LARM DELAY
?? to Choose, ENTER to
Select, MENU to Exit
A
larm Level: xxx Seconds
Pressure Sensor Auto-Zero Note: Every time blower is shut off, wait 100 seconds and thentake zero data for 30 seconds.
A
uto- Zero occurs while the blower is off for the entire time span. The “span” or difference between zero airflow and 70
CFM
is saved from “Pressure Sensor” calibration in step 2. Also, flash icon on LCD to shut Auto-Zero.
A
IRFLOW SENSOR
A
UTO
-
ZERO
1. Disable
2. Enable
A
IRFLOW SENSOR ZERO
Please Wait:
150
Enable
BOX TYPE
1. Positive Pressure
2. Negative Pressure
PRESSURE SENSOR
LEVEL ADJUST
Wait 150 Seconds