Manual

Chapter 2: Prerequisites
Product Service 1-800-522-7658
11
PROTECTOR STAINLESS STEEL ULPA-FILTERED GLOVE BOX (230V)
Static Pressure Test Results
0.30
0.51
0.72
0.83
1.00
0.52
0.89
1.23
1.41
1.67
0
0.5
1
1.5
2
2.5
20 25 30 35 40 45 50 55 60
Airflow Volume
(Cubic Feet per Minute)
Negative Static Pressure
(Inches of Water Column)
Static Pressure Inside Glove Box
Total Static Pressure at Blower
Intake (Suction Side)
Static Pressure test results on the Protector ULPA-Filtered Glove Box
Note: Test results are approximately the same for positive pressure application
PROTECTOR ULPA-FILTERED GLOVE BOX (230V)