Protector CA (Fiberglass) Oxygen Removal with AtmosPure Re-Gen Gas Purifier
7.9
969.2
63.9
28.0
15.3
4.3
2.4
1.6
1.2
1.0
0.9
0.77
0.1
1
10
100
1000
0 20 40 60 80 100 120 140 160
Time (Minutes)
Oxygen (PPM)
This chart shows oxygen level decrease from 969.2 ppm to 0.77 ppm after
150 minutes. Two separate calibrated oxygen monitors confirmed results.