Instruction Manual
Chapter 5: Using Your Glove Box
Product Service 1-800-522-7658
44
Auto Pressure Controller Specifications
Glove Box
Pressure Control
Operating Range Default Settings
Pressure Control Limits -5 inches of water column to
+5 inches of water column;
[+/- 12.6 mBar]
+/- 5 inches of water column;
[+/- 12.6 mBar].
Manual Control Available Purge (Gas out);
Fill (Gas in)
Audio Alarm level Whenever pressure
outside limits
Glove Box Purge/Fill
Purge/Fill Pressure Limit +/- 5 inches of water column;
[+/- 12.6 mBar].
+/- 5 inches of water column;
[+/- 12.6 mBar].
Manual Control Available Purge (Gas out);
Fill (Gas in)
Number of Purge/Fill
Cycles
0 to 499 – main chamber 1
Audio Alarm level Whenever pressure
outside limits
Transfer Chamber
Purge/Fill
Purge Pressure Limit -29.5 inches of Mercury/Hg
[-999 mBars]
-29.5 inches of Mercury/Hg
[-999 mBars]
Fill Pressure Limit 0 inches of Mercury/Hg
[0 mBars](atmosphere)
0 inches of Mercury/Hg
[0 mBars](atmosphere)
Number of Purge/Fill
Cycles
0 to 199 – transfer chamber 1
Audio Alarm level -29.5 inches of Mercury/Hg
[-999 mBars]
Whenever pressure/vacuum
outside limits