The common service operations necessary to maintain the Evidence Drying
cabinet for peak performance are listed below.
Routine Maintenance Schedule
Weekly
Using 70% ethanol, or a suitable disinfectant, surface disinfect the
inside of the cabinet.
Using an appropriate glass cleaner, or LabSolutions Glass & Surface
Wipes, Labconco part # 1570000; clean the door and the surface of
the UV lamp.
Operate the cabinet blower, noting if the Filter System Light turns on.
Ensure the cabinet water shut off valve is in the off position.
Monthly (or more often as required)
Using a damp cloth, or LabSolutions Glass & Surface Wipes,
Labconco part # 1570000; clean the exterior surfaces of the cabinet,
particularly the front and top of the cabinet, to remove any
accumulated dust.
All weekly activities.
Check the prefilters; replace both if needed.
Annually
All monthly activities.
Replace HEPA and granular activated carbon filters.