Chapter 5: Using Your Glove Box
Product Service 1-800-522-7658
22
Loading Materials and Equipment (see Figure 5-1)
Load only the materials required for the procedure. Do not overload
the glove box.
Be aware that gloves will inflate when the glove box is operated
under negative pressure. Ensure items in the glove box will not
obstruct the movement of the gloves.
After loading, wait 2-5 minutes at 50-100% blower speed to purge
airborne contaminants from the work area before beginning
procedures.
Work Techniques
Segregate all clean and contaminated materials in the work area.
Final Purging (see Figure 5-1)
All open trays, weigh vessels or containers must be sealed before
being removed from the glove box.
Objects in contact with contaminated material must be surface
decontaminated before removal from the glove box.
The interior surfaces should be cleaned per user’s protocol.
Upon completion of work, the glove box should be allowed to operate
for 5 minutes undisturbed at 100% speed or 10 minutes at 50% speed
to purge and dilute airborne contaminants from the work area.
After purging the main chamber, the transfer chamber should also be
purged for 5 minutes at 100% speed or 10 minutes at 50% speed with
the inner door open, before removing sealed material.
Shutdown
Turn off the exhaust system and light.
Keep outer door closed when not in use.