User guide

Labconco Instruction Sheet 1049200, Rev -, ECO H250 01/13 2 of 5
Set-up
1. The Small Stoppering Chamber has a stoppering travel distance of ½”(12.7mm), it can
accommodate vials that range in height from 1.38”(35mm) to 2.48”(63mm) tall by using
zero, one, or two of the spacers which are provided. Labconco offers 6 sizes of serum
vials that can be used in the Small Stoppering Chamber. See the chart below to determine
the number of spacers required for specific vials and the total number of vials the
chamber will hold of each size.
Size
Part no.
Number of
Bottles/run
Spacers
Required
Vial Dimensions
Diameter x Height
Split Stopper
Part No.
Part No.
2 ml
7575000
48
2
18mm x 35mm
7576000
3ml
7575200
43
2
17mm x 28mm
7576000
5 ml
7573000
21
1
23mm x 47mm
7576200
10 ml
7573200
19
1
25mm x 54mm
7576200
20 ml
7573400
11
0
32mm x 58mm
7576200
30 ml
7573600
8
0
37mm x 63mm
7676200
2. Once the vial size is selected place a stoppered vial in the chamber with the tray and
correct number of spacers. The top of the stopper should be located between the two
scribed lines marked on the inside diameter of the clear chamber. If the vial is lower than
the bottom line, the Stoppering device will not stopper fully. If the vial is higher than the
top line the lid will prematurely stopper the vials when it is placed on top of the chamber.
Caution: only one size (height) vial can be Lyophilized and stoppered during the same
run.
Figure 2