1.2 Events / actions
Safe airflow
Airflow above alarm level ( e.g. > 60fpm )
Green LED on
Low airflow
Airflow below alarm level
Red LED on ( Flashing )
Audible alarm sounds -- can be muted via Mute pushbutton
Low air relay operates
Reset : -- when airflow rises above Low air level the Low air alarm resets automatically
Sash High
When the input configured as Sash High is activated
Amber LED on
Audible alarm sounds
Audible can be muted via Enter pushbutton
Reset: -- when Sash lowered to safe position and input de-activated.
Setback
When Night set-back input is activated
Red LED on ( Flashing)
Audible alarm muted
External alarm
When External alarm input is activated
Red LED on ( Flashing)
Audible alarm sounds – can be muted via Enter pushbutton