Owner manual
3
NOT ACTIVE Not active
ACTIVE ON CLOSING CONTACT
ON INPUT 2
ACTIVE ON OPENING CONTACT
NONE None
ALARM DISABLE
NIGHT SET-BACK
EXTERNAL ALARM
SASH HIGH
ON INPUT 2 - FUNCTION
HIGH/LOW
NOT ACTIVE
ACTIVE ON CLOSING CONTACT ACTIVE ON CLOSING CONTACT
ON INPUT 3
ACTIVE ON OPENING CONTACT
NONE
ALARM DISABLE
NIGHT SET-BACK
EXTERNAL ALARM
SASH HIGH SASH HIGH
ON INPUT 3 - FUNCTION
HIGH/LOW
CONTACT CLOSES ON ACTIVATION Contact closes on activation
ON OUTPUT RELAY 1 ACTIVATION
CONTACT OPENS ON ACTIVATION
CONTACT CLOSES ON ACTIVATION Contact closes on activation
ON OUTPUT RELAY 2 ACTIVATION
CONTACT OPENS ON ACTIVATION
CONTACT CLOSES ON ACTIVATION Contact closes on activation
ON OUTPUT RELAY 3 ACTIVATION
CONTACT OPENS ON ACTIVATION
NONE
OUTPUT RELAY 1
OUTPUT RELAY 2
ON LOW AIR ALARM - RELAY
OUTPUT RELAY 3
OUTPUT RELAY 3
NONE None
OUTPUT RELAY 1
OUTPUT RELAY 2
ON DISABLE ALARM - RELAY
OUTPUT RELAY 3
ON ON
ON SASH HIGH REPEAT TIMER
OFF
ON SASH REPEAT TIME MIN 5
NONE None
OUTPUT RELAY 1
OUTPUT RELAY 2
ON SASH HIGH RELAY
OUTPUT RELAY 3
NONE None
OUTPUT RELAY 1
OUTPUT RELAY 2
ON HIGH / LOW RELAY
OUTPUT RELAY 3
MAINTAIN LOW AIR ALARM Maintain Low Air alarm
ON NIGHT SET-BACK
REDUCE LOW AIR ALARM
ON NIGHT SET BACK REDUCED ALARM FPM 49
NONE None
OUTPUT RELAY 1
OUTPUT RELAY 2
ON NIGHT SET-BACK RELAY
OUTPUT RELAY 3