Owner manual
2
FILE NUMBER
AFA 1000 / 1 -- ALARMS -- ( Velocity )
TYPE
CUSTOMER
FACTORY DEFAULTS
The AFA1000 alarm has been factory set to the listed parameter settings. To change
any parameter setting press and hold the enter button until the "Main Menu" screen is
displayed, select "Configure" to view the parameters and use the +/- and Enter buttons
to change the settings.
FOR FULL INSTALLATION / CALIBRATION / CONFIGURATION DETAILS PLEASE
REFER TO THE DATA BOOKLET PROVIDED.
Configuration Menu: -
Menu
Calibration Config
Units Default Settings
M/SEC
ON DISPLAY UNITS
FPM
FPM
ON LOW AIR ALARM FPM 60
OFF
ON LOW AIR CUTOFF
ON
ON
ON CUTOFF SPEED FPM 30
WARNING AIR ALARM FPM 63
ON
ALARM TO WARNING AIR HYSTERESIS FPM
3
YES
ON HIGH AIR ALARM
NO
OFF
ON HIGH AIR ALARM FPM 295
ON LOWER AIR SAMPLE FLUCTUATIONS % 3
ON HIGHER AIR SAMPLE FLUCTUATIONS % 3
ON LOWER/HIGHER AIR SAMPLE DIFFERENCE FPM 20
ON WARN TO ALARM AIR TIME SECONDS 5
ON ALARM TO WARN AIR TIME SECONDS 1
YES ON
ON SHOW AIR FLOW
NO
YES
ON SHOW TIME LINE
NO = SHOW BAR GRAPH
OFF
ENABLED ENABLED
ON AUDIBLE ALARM
NOT ENABLED
ON SENSOR DIFFERENCE
%
10
Configuration Menu
NOT ACTIVE
ACTIVE ON CLOSING CONTACT ACTIVE ON CLOSING CONTACT
ON INPUT 1
ACTIVE ON OPENING CONTACT
NONE
ALARM DISABLE
NIGHT SET-BACK NIGHT SET-BACK
EXTERNAL ALARM
SASH HIGH
ON INPUT 1 - FUNCTION
HIGH/LOW