SPECIFICATIONS
21
Dimensions: 47 cm w x 50 cm d x 8.9 cm h
(18-1/2" w x 19-1/2" d x 3-1/2" h)
Gel Size: 33.6 cm x 42.3 cm maximum
(13.2 in. x 16.7 in. maximum)
Temperature Range: Continuous and adjustable from 40°C –80°C
Power Consumption: 115 VAC - 16 Amps
(without vacuum pump) 230 VAC - 8 Amps
Recommended Minimum 50 liters/minute free air displacement
Vacuum : 25 in. Hg. minimum vacuum
Vacuum Port: l/4 inch I.D.
Vacuum Port Switch 115 VAC - 10 Amps maximum
Rating: 230 VAC - 5 Amps maximum
Environmental Conditions
This equipment is designed to be safe under the following conditions:
Indoor use.
Altitude up to 6562 Ft. (2000m).
Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing
linearly to 50% relative humidity at 104° F (40°C).
Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
Transient overvoltages according to Installation Categories II (Overvoltage
Categories per IEC 1010).
Pollution degrees 2 (Normally only non-conductive foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface
resistivity occurs. Occasionally, however, a temporary conductivity caused by
condensation must be expected), in accordance with IEC 664.