Refer to the following if your CentriVap fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance. The following failure codes may appear on the display
when problems are sensed by the internal self check system.
CAUTION: Disconnect power before servicing the CentriVap.
DISPLAY ERROR CODE CAUSE CORRECTIVE ACTION
Heat Sensor
Sensor failure Replace sensor assembly.
Connection failure Repair connection.
Close Lid
Lid open Close lid.
Latch Fail
Solenoid failure Check component.
Switch or sensor failure Check connections.
Mem Fail P1
Memory failure Push program button #1.
New memory IC chip Push program button #1.
Bad memory IC chip Call Labconco – Replace IC chip or control PCB.
MOTOR ERROR
Defective motor Replace motor.
Hall effect sensor failure Replace sensor.
Wire failure Replace wire.