Owner manual
Appendix C: CentriVap Specifications
Product Service 1-800-522-7658
42
Water
bp 100
Tube Size
(ml)
Number
of
Samples
Sample
Size (ml)
Heater
Temp
(C)
Heat
Boost
Vacuum
Pump
Time to
Dry
(min)
Cold
Trap
Rate
Overall
(ml/min)
50 12 25 45 off Diaphragm 1397 no 0.21
50 12 25 100 off Diaphragm 445 no 0.67
50 12 25 100 on Diaphragm 430 no 0.70
15 18 10 45 off Diaphragm 841 no 0.19
1.5 132 1 45 off Diaphragm 427 no 0.30
1.5 132 1 60 off Diaphragm 303 no 0.44
1.5 132 1 75 off Diaphragm 196 no 0.67
1.5 132 1 100 off Diaphragm 117 no 1.13
1.5 132 1 100 on Diaphragm 108 no 1.22
50 12 25 45 off rotary vane 1002 yes 0.30
50 12 25 100 off rotary vane 424 yes 0.70
50 12 25 100 on rotary vane 373 yes 0.80
15 18 10 45 off rotary vane 565 yes 0.32
15 18 10 60 off rotary vane 456 yes 0.39
1.5 132 1 45 off rotary vane 299 yes 0.44
1.5 132 1 75 off rotary vane 207 yes 0.64
1.5 132 1 100 off rotary vane 131 yes 1.01
1.5 132 1 100 on rotary vane 122 yes 1.08
Vacuum Pumps:
Diaphragm – Rated at 12 mbar ultimate vacuum and 34 L/min displacement
Rotary vane – Rated at 2.0 x 10
-3
mbar ultimate vacuum and 195 L/min displacement
Chamber preheated prior to each run to run temp.