Safety Cabinets 34610 Series Instruction Manual

Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
40
Start-up
Turn off UV light if included.
Slowly raise the sash until the bottom of the sash aligns with the sash
indicator decal located on the left side of the work area.
Turn on the fluorescent light and cabinet blower if the Smart-Start
features have not been activated.
Check the air grilles for obstructions, and note the filter gauge
reading.
Allow the cabinet to operate unobstructed for 5 minutes.
Wash hands and arms thoroughly with germicidal soap.
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if
appropriate.
Wipe-Down
Raise the sash to its full open position (approximately 21.75 inches or
552 mm). Mute the alarm by depressing the "Alarm Silence" switch.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload
the cabinet.
Do not obstruct the front, side, or rear return air grilles.
Large objects should not be placed close together.
Slowly close the sash until it is in the correct operating position.
After loading the cabinet, wait two to three minutes to purge airborne
contaminants from the work area.
Work Techniques
Keep all materials at least 4 inches (100 mm) inside from the sash,
and perform all contaminated operations as far to the rear of the work
area as possible.
Segregate all clean and contaminated materials in the work area.