Datasheet

For inquires or orders :
Safety Warnings :
Please read the "Safety Warnings" in the instruction manual supplied with the instrument thoroughly and completely
for correct use. Failure to follow the safety rules can cause fire, trouble, electrical shock, etc. Therefore, make sure
to operate the instrument on a correct power supply and voltage rating marked on each instrument.
The contents of this leaflet are subject to change without notice.
In consideration of the environment, soy ink and recycled paper were used in this publication.
KEW 2500-1E Mar. 12 AD
MODEL 7256
Output cord
Recorder or Logger
measured values (mA) % values
MODEL7256
Output cord(Optional)
DC10mV/DC1mA
φ
6mm
Optional Accessory
Included Accessories
KEW 2500 Specifications
Spotlight LED & Backlight Display
Applicable to signal measurement
on process and building instrumentation
Digital Multimeter
Analog output terminal for recorder connection
Applications
Dual display gives simultaneous readout of % values and
measured values(mA)
DC A 20/100mA(Auto ranging)
±
0.2%rdg
±
5dgt(0.00mA
21.49mA)
±
1.0%rdg
±
5dgt(21.0mA
120.0mA)
Conductor size
φ
6mm max.
Analog output Recorder: DC1000mV against DC100mA
Withstand voltage 2210V AC for 5 seconds
Applicable standards IEC 61010-1, 61010-2-030 CAT.
300V
IEC 61010-2-032
IEC 61326-1, 61326-2-2
IEC 60529 IP40
Operating temperature & humidity -10
+50
< 85%
Storage temperature & humidity -20
+60
< 85%
Power sour
ce R6/LR6(AA) (1.5V)
×
4
Dimensions 111(L)
×
61(W)
×
40(D)mm : Display unit
104(L)
×
34(W)
×
20(D)mm : Sensor
700mm : Sensor cable
Weight Approx. 290g (including batteries)
Accessories 9096(Carrying case) LR6(AA)
×
4
Instruction manual, Calibration certificate
Optional 7256(Output cord)
Fluorescent buttons to increase
operating in low light conditions
Diameter of measurable
conductor :
φ
6mm max
Measures DC mA signals used in industrial automation equipment, PLC, transmitters,
process control and analog I/0 systems without breaking the loop.
Kew 2500 is ideal for measuring the level of 4-20 mA signals correlating process indi-
cation with real physical value.
Very useful to maintain and troubleshoot process and automation equipment without
breaking the loop and stopping the process.
Note: The Auto-power OFF can be disabled for long recording