User Guide

Kunak AIR Pro QUICK START GUIDE v.6
P. E. La Muga 9, 4º Ofic. 1, Orcoyen, 31160, Navarra, Spain +34 848 470 055 info@kunak.es www.kunak.es/en 7
Install the device far away from very close pollution sources and without enough air flow
(chimneys, air conditioners, etc), that may disturb your readings. If a sonometer is also
going to be placed, install it in a place with low wind impact, avoiding vibration sources and
avoiding electromagnetic sources.
Install the unit between 3 and 4 meters high.
If the unit is installed with solar panel, locate the devices far away from buildings, trees,
etc…. to be sure that the solar panel receives maximum solar radiation to guarantee a
correct battery recharge. Otherwise, Kunak AIR Pro will run out of battery.
if the device is intalled with a solar panel in the North hemisphere, the solar panel must
face to the South, and to the North in case of the South hemisphere.
1.10 Installation to Mast or Wall Mount
The Kunak AIR Pro device must be mounted in a determined way due to some design restrictions.
The fixed Kunak AIR Pro model should be mounted with the five cartridge sensor holes facing down and the
Kunak AIR Pro sticker facing to the front.
The steps to install the Kunak AIR Pro to a wall or mast are described below. The device is delivered
completely mounted into the solar shield.
Different types of possible installations: vertical mast, horizontal mast or wall. The minimum mast diameter
on which the device can be hung is 3cm.
Video tutorial of the steps described below
Vertical mast
Horizontal mast
Wall
Figure 4 Types of possible installations