User Guide

Kunak AIR Pro QUICK START GUIDE v.6
P. E. La Muga 9, 4º Ofic. 1, Orcoyen, 31160, Navarra, Spain +34 848 470 055 info@kunak.es www.kunak.es/en 4
1.5 LEDs Indicators and Display Buttons
The LEDs below the display indicate the current status of the device.
RED LED
GREEN
LED
INFORMATION
ON
ON
Powering ON or OFF
ON
OFF
Error status
OFF
ON
Measuring or connecting to the Kunak Cloud
ON (blink 1Hz)
OFF
Device ON at stand-by mode and charging
OFF
OFF
Between measures or powered off
There are two buttons below the LCD that will change the LCD messages. To access the functionality of the
buttons, for the upper symbols briefly press the button, and for the lower symbols, press the button for 5
seconds.
Icon
Name
Hold
Description
Info
Simple
Show information about the state of the device and
the sensors
Alarms
5 sec
Currently not available
Maintenance
Simple
Currently not available
Sleep
5 sec
Switching the working mode to stand-by or to
normal mode..
1.6 Installation of the Smart Gas Cartridge
The cartridge can be installed at any slot of the Kunak AIR Pro device (always with the device switched off).
When is turned ON, the device will automatically detect the cartridge and its position showing them at the
LCD display. The positions are:
Figure 2 Cartridges position
1.7 Pluggable Probes or Extra Sensors
There are 5 plugs at the bottom of the Kunak AIR Pro device to connect the power and extra sensors. Each
M12 plug is different and only fits its specific connector.
The VOCs sensor cartridge ionises ambient air generating small amounts of O3, so avoid
installing it near O3, NO2 or SO2 sensors. For example, place the VOCs sensor in slot 5 and
the O3, NO2 and SO2 sensors in slots 1, 2 and 4; avoid placing them in slot 3.