Owner's manual
Handbook BM 100 A 69
F8 – Markers Menu
Displays a real-time evolving graph of error markers which are either displayed as active (value 1) or
inactive (value 0).
Item Item Description
(grey – error marker)
Notes
1 No reference as marker 1 on gauge display
2
No level as marker 2 on gauge display
3
Level frozen as marker 3 on gauge display
4
No interface as marker 4 on gauge display
5 Interface frozen as marker 5 on gauge display
6
Communication as marker 6 on gauge display
7
No reference error
8 No level error
9
No interface error
10
No EOS error EOS=End Of Scan signal
11 Dead zone error
12
Microwave error
13 Slider control to move from to the next time slice Window function
14 Date and time of time slice. Window data
15 Time slice indicator on graph Window data
16 Error marker “off” Window data
17 Error marker “on” Window data
18 Slider to scroll display to other error markers Window function, see below
19 Error marker status Window data “0” = off ,“1” = on