Owner's manual

Handbook BM 100 A 29
How to protect gauge configuration
User menu functions 1.4.2 Entry Code 1 & 1.4.3 Code 1
These functions are used to activate the restricted-access code and lock the settings: these will be
impossible to modify without entering the correct code. Do not lose this code!
Example procedure 3A:
to activate access code function,
to enter new code and exit configuration mode.
Step Action Press buttons to
finish step
Information displayed at the end
of each step
1 Start from Operating Mode
screen
n/a 6750
LEVEL mm
2 Go to function (fct.) 1.4.2
→,→,↑,↑,↑,→,→
Fct. 1.4.2
ENTRY.CODE1
3 Enter function data set field
(default value: No)
No
4 Select Yes to confirm activation
of access
code.
Yes
4 Exit function and go to fct.
1.4.3.
↵,↑
Fct. 1.4.3
CODE 1
CodE1
---------
5
Enter function 1.4.3 data set
field.
Default: (
↑↑↑↵↵↵→→→
)
6 Enter new 9-keystroke code
↑,↑,↑,↑,↑,↑,↑,↑,↑
CodE1
---------
7 Re-enter new 9-keystroke code
to confirm. Automatic exit from
data set field to function.
↑,↑,↑,↑,↑,↑,↑,↑,↑
Fct. 1.4.3
CODE 1
8 Exit function, submenu and
menu. Save new
configuration and return to
Operating Mode.
↵,↵,↵,
6750
LEVEL mm
Example procedure 3B:
to re-enter the configuration mode from the Operating mode with the restricted-access
function activated.
9 Start from Operating Mode
screen
n/a 6750
LEVEL mm
10 Press RIGHT button to enter
Configuration Mode. Screen
displays access code entry.
CodE 1
---------
11 Type 9-keystroke access code.
User menus are now
accessible. A typing error
(impossible to delete incorrect
keystroke) will quit the user
back to the operating mode.
↑,↑,↑,↑,↑,↑,↑,↑,↑
Fct. 1.0.0
OPERATION