Owner's manual

10 Handbook BM 100 A
1 Agitator
2 Support beam perpendicular to the pulse
direction
3 Abrupt changes in tank cross section
4 Heating tubes
5 Alternative solution: reference chamber -
electromagnetic field is contained within
chamber.
6 Gauge electromagnetic field :
Any intruding metallic object will be
detected in this zone if perpendicular to
the emitted pulse direction.
= Do not fit the process connection near to these objects.
Probe Type Recommended minimum distance of probe from
objects inside the tank in millimetres (inches)
Single (types F, H and K) 300 (12)
Twin (types A, B, G and L) 100 (4)
Coaxial (type D) 0 (0)
Electromagnetic field shape around probe, by type (not to scale)
Single (types F, H and K) Twin (types A, B, G and L) Coaxial (type D)
No beam angle for any probe type.