Installation and operating instructions BM 100 A 17
2.2.2 Wiring connections
Open the signal converter housing rear cover, using the plastic wrench supplied. The terminal
connections are labelled. The standard connections are shown below:
Before starting to wire:
check that the power supply corresponds to the power board installed and
check which output option you have selected.
This information will be indicated on the underside of the rear housing cover and on the gauge
nameplate.
Terminal layout: non – Ex version
X = Terminal not used (X) = Terminal not used except for RS485 outputs
Ensuring a good contact and protection of wire strands
Local regulations concerning electrical wiring must be followed and obeyed. If no details are
given, we recommend :
crimped metal sheaths over the wire strands
power supply cables should be rated for 500 V at least, with a cable diameter of 0.5 to 1.5 mm
/ 0.02” to 0.06” (non-Ex applications only).
the output current cable diameter should be from 0.5 to 0.75 mm / 0.02” to 0.03”
PE Ground Terminal notes
The internal earth connection shall be used according to local applicable installation standards, in
Europe the Low voltage Standard prescribes the connection of the yellow/green cable in case of
230VAC.