Quickstart EN User guide

2 INSTALLATION
6
BATCHFLUX 5500 C
www.krohne.com 05/2014 - 4003386301 - QS BATCHFLUX 5500 R03 en
2.4 Storage
Store the device in a dry and dust-free location.
Avoid lasting direct exposure to the sun.
Store the device in its original packaging.
Storage temperature: -50 ...+70°C / -58...+158°F
2.5 Pre-installation requirements
Make sure that you have all necessary tools available:
Small wrench (M5) for connection to ground
Torque wrench for installing flowmeter in pipeline
Assecoires necessary for the correct installations are available on request at the manufacturer
Make sure that these accesoires are available before starting installation;
O- ring / L-ring gasket
Special pipe flanges
Stud bolt with lockwasher, plain washer and nut
2.6 General requirements
INFORMATION!
To facilitate servicing and/or exchanging of the device, please note that:
it must be possible to shut off the flow through the pipeline ( control valve upstream in pipeline).
Drain the pipeline before removing device ( provide drain valve)
INFORMATION!
The following precautions must be taken to ensure reliable installation.
Make sure that there is adequate space to the sides.
Protect the signal converter from direct sunlight and install a sun shade if necessary.
Support the pipeline on both side of the flowmeter.
Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 60068-2-64.