7 Functional checks
7.1 Zero check with signal converter IFC 015
Please note!
Do not open the housing of the
BATCHFLUX IFM 5015 K.
Danger of contamination with substances likely
to destroy the moisture barrier of the electronic
equipment (e.g. if CIP or SIP cleaned from the
outside).
Therefore, please contact your KROHNE
Service engineer before you open the housing.
Switch off
power source
before o
pening the housing
Set “zero“ flow in the pipeline, but make
sure that the measuring tube is
completely filled with fluid.
Switch on the system and wait 15 minutes.
Press the following keys for zero
measurement
Key Displayed Description
If “YES“ is set under Fct. 3.4 ENTRY CODE, key-in the 9-
keystroke CODE 1:
Fct. 1.0 OPERATION
2x
Fct. 3.0 INSTALL.
Fct. 3.1 LANGUAGE
2x
Fct. 3.3 ZERO SET
CALIB. NO
CALIB. YES
0.00 - - - - - / - - - Flowrate set in displayed unit, see Fct. 1.4 DISPLAY,
subfunction “DISP.FLOW“.
Zero measurement in progress, duration approx. 10
seconds.
When flow “> 0“, “WARNING“ notice appears, confirm
with key
.
STORE NO
If new value not to be stored, press key (3x) 4x =
return to measuring mode.
STORE YES
Fct. 3.3 ZERO SET Store new zero value.
(2x) 3x
- - - - - - - - - - - - / - - - Measuring mode with new zero.