Manual

Installation and operating instructions 32 BATCHFLUX
5.8 Status output S (option)
No HARDWARE!
Function must be set to “OFF”.
Fct. 1.7 IND. OUTP. S
Press key
.
Select function for the status output,
press ke
y .
OFF (switched off,
no function)
ON (indicates that
flowmeter is operative)
V/R INDIC. (indicates direction for
current output and
pulse output, F/R mode)
TRIP.POINT (setting range: 002 -
115 PERCENT of Q
100%
,
full-scale range)
endwert)
ALL ERROR (indicates all errors)
FATAL.ERROR (indicates
Fatal Error only)
Transfer to number setting with key
, 1st
digit (cursor) flashes. Change flashing digit
(cursor) with keys
and , shift cursor 1
place to right or left with keys
and .
Press key
to return to
Fct. 1.7 IND. OUTP. S.
Characteristics of
Switch open Switch closed
status output
OFF
no function
ON (e.g. operation indicator)
Power OFF Power ON
F/R INDIC. Forward flow Reverse flow
TRIP.POINT
Inactive Active
ALL ERROR
Error(s) No error(s)
FATAL. ERROR
Error No error
For factory settings, please refer to the report on settings and to Sect. 3.2.
Connection diagrams: see Sect. 2.2.
5.9 Language
Fct. 3.1 LANGUAGE
Press key
.
Select lan
guage for texts in display
D (German)
GB (English)
F (French)
others available on request
Select with keys
and .
Press key
to return to Fct. 3.1 LANGUAGE.
5