Manual

Installation and operating instructions 30 BATCHFLUX
5.6 Current output I
No HARDWARE!
Function must be set to “OFF”.
Fct. 1.5 CUR.OUTP. I (not available, no
HARDWARE !)
Press key
.
FUNCTION I = Select function for the
current output, press key
OFF (switched off,
do not change setting,
no hardware available)
1 DIR. (1 flow direction)
2 DIR. (2 flow directions, F/R mode,
forward/reverse)
Select with keys and .
Transfer to subfunction “RANGE I“
with key
. Exception: when “OFF“ selected, return to
Fct. 1.5 CUR.OUTP. I.
RANGE I = Select measuring range,
press ke
y
0 - 20 mA
4 - 20 mA
Select with keys and .
Transfer to subfunction “I ERROR“ with key
.
I ERROR = Set error value,
press key
0 mA
3.6 mA (only possible when
range “4 - 20 mA“
selected)
22 mA
Select with keys and .
Press key
to return to Fct. 1.5 CUR.OUTP.
I.
Please refer to Sect. 3.2
“Factory settings“.
Refer to Sect. 2.3 for connection diagrams,
and to Sect. 5.14 for characteristics.
5.7 Pulse output P
Fct. 1.6 PULS.OUTP. P
Press key
.
FUNCTION P = Select function for the
pulse output, press key
OFF (switched off, no function)
1 DIR. (1 flow direction)
2 DIR. (2 flow directions, F/R mode,
forward/reverse)
Select with keys and .
Transfer to subfunction “SELECT P“
with key
. Exception: when “OFF“ selected, return to
Fct. 1.6 PULS.OUTP. P.
SELECT P = Select pulse type,
press ke
y
100-1000 Hz
PULSE/VOL. (pulses per unit volume,
flow)
PULSE/TIME (pulses per unit time
for 100 % flow)
Select with keys and .
Transfer to subfunction “PULSEWIDTH“
with key
.
Note: when 100-1000 Hz selected,
return to Fct. 1.6 PULS.OUTP. P.
PULSEWIDTH = Select pulse width,
press ke
y
50 mSec F
max
= 10 Hz
100 mSec F
max
= 5 Hz
200 mSec F
max
= 2.5 Hz
500 mSec F
max
= 1 Hz
for all F
min
= 0.0056 Hz
(= 20 pulses / hr)
Select with keys
and .
Transfer to subfunction “VALUE P“ with key
or return to Fct. 1.6 PULS.OUTP. P,
depending on selected pulse type in
subfunction “SELECT P“.
5