Manual
Installation and operating instructions
23
BATCHCONTROL
6.3 Signal converter
Power supply
Voltage 24 V DC, ± 20%
Power consumption ≤ 5 W excl. external loads
Electrical connection two M12 plug-in connector
Operator control All operating data factory-set to your specifications.
Available as option for change of operating data:
- KROHNE software for operator control via PC.
Options connected to the CAN bus interface
6 Valve outputs low or high side activ
Function programmable to control valves,
short and cable broken detect
Voltage 24 V DC, ± 20%
Load rating I
max
≤ 200 mA
Load short I
short
≤ 2 A
6 Analogue inputs alternative to the outputs
Function programmable for alarm messages,
to control the filling process
Voltage 0 – 11 V DC, resolution of 8 bit
Impedance input 1 and 2: 220 kOhm
Input 3 to 6: 22 kOhm
Electronic temperature sensor
Function programmable for alarm messages
Range -20°C … +100°C
Error ± 3 K
Fluid temperature sensor
Function programmable for alarm messages
Range -20°C … +140°C
Error ± 3 K
CAN bus
Baud rate 20k … 1M baud
6