3 Start-up
Before powering the system, check that it has been installed correctly according to Sections 1 and
2.
The compact flowmeter is delivered ready for operational use. All operating data have been factory
set in accordance with your specifications.
Power the unit, and the flowmeter will start process flow measurement immediately.
3.1 Check for availability
Please note!
Do not open the housing of the BATCHCONTROL IFM 5014C.
Danger of contamination with substances likely to destroy the moisture barrier of the electronic
equipment (e.g. if CIP or SIP cleaned from the outside).
Therefore, please contact your KROHNE Service engineer before you open the housing.
The measurement status is signalled by the light-emitting diodes (LED) below the cover of the
converter housing (see amplifier board on right).
LED 1 LED 2 Function
flashing off in order
flashing flashing overdriving
on flashing Fatal Error
(defective operating
parameter)
off on defective hardware
off off no supply voltage
or hardware
is defective
All operating data can be set and stored by means of a personal computer via the CAN bus
interface. The digital CAN bus interface allows the complete filling procedure to be graphically
represented on the PC, thus providing visualization of system and valve properties.
3.2 Factory settings
All operating data are factory-set according to your order specifications.
The table of standard factory settings is shown at the end of the manual.
3
LED 1
LED 2