ALTOSONIC V
ModBus Manual 0300 rev07 E 7.30855.35.00 Page 47 of 64
30 59 F32+RW
UFP batch control:
Normal:
Setup=9 (if UFP batch1 status batch=0 is no batch)
Cancel=5 (if UFP batch1 status batch=1 is set-up)
Start batch=119 (if UFP batch1 status batch=1 is set-up)
End batch=229 (if UFP batch1 status batch=1 is running)
reset print=1009 (if UFP batch1 status batch=5…10 printing)
Confirm ticket=779 (if UFP batch1 status batch=10 is confirm)
Continuous pipe line measurement, ticket on demand:
End with no reset values=559 (if UFP batch1 status batch is not
printing)
End with reset values=229 (if UFP batch1 status batch is not
printing)
Reset printing=1009 (if UFP batch1 status batch 5…10
is printing)
559 and 229 have automatic freeze of batch values (on Modbus)
for a maximum of 30 seconds (see also Boolean 2075)
6 2
5…1009
On command value input:
Return 0 if not accepted
Return 1 if accepted
Reset to -99999 after 5 seconds
31 61 F32+R Solartron1: time to update a parameter (read only) 5 2 s, max 30 sec
32 63 F32+RW Solartron1 K0 6 2 device calib. Param.
33 65 F32+RW Solartron1 K1 6 2 device calib. Param.
34 67 F32+RW Solartron1 K2 6 2 device calib. Param.
35 69 F32+RW Solartron1 K18 6 2 device calib. Param.
36 71 F32+RW Solartron1 K19 6 2 device calib. Param.
37 73 F32+RW Solartron1 K20A 6 2 device calib. Param.
38 75 F32+RW Solartron1 K20B 6 2 device calib. Param.
39 77 F32+RW Solartron1 K21A 6 2 device calib. Param.
40 79 F32+RW Solartron1 K21B 6 2 device calib. Param.
41 81 F32+R Solartron2: time to update a parameter (read only) 5 2 s, max 30 sec
42 83 F32+RW Solartron2 K0 6 2 device calib. Param.
43 85 F32+RW Solartron2 K1 6 2 device calib. Param.
44 87 F32+RW Solartron2 K2 6 2 device calib. Param.
45 89 F32+RW Solartron2 K18 6 2 device calib. Param.
46 91 F32+RW Solartron2 K19 6 2 device calib. Param.
47 93 F32+RW Solartron2 K20A 6 2 device calib. Param.
48 95 F32+RW Solartron2 K20B 6 2 device calib. Param.
49 97 F32+RW Solartron2 K21A 6 2 device calib. Param.
50 99 F32+RW Solartron2 K21B 6 2 device calib. Param.
51 101 F32+R Sarasota1: time to update a parameter (read only) 5 2 s, max 30 sec
52 103 F32+RW Sarasota1 K 6 2 device calib. Param.
53 105 F32+RW Sarasota1 T0 6 2 device calib. Param.
54 107 F32+RW Sarasota1 D0 6 2 device calib. Param.
55 109 F32+RW Sarasota1 Nt 6 2 device calib. Param.
56 111 F32+RW Sarasota1 Np 6 2 device calib. Param.
57 113 F32+RW Sarasota1 Tcal 6 2 device calib. Param.
58 115 F32+RW Sarasota1 Pcal 6 2 device calib. Param.
59 117 F32+R Sarasota2: time to update a parameter (read only) 5 2 s, max 30 sec
60 119 F32+RW Sarasota2 K 6 2 device calib. Param.
61 121 F32+RW Sarasota2 T0 6 2 device calib. Param.
62 123 F32+RW Sarasota2 D0 6 2 device calib. Param.
63 125 F32+RW Sarasota2 Nt 6 2 device calib. Param.
64 127 F32+RW Sarasota2 Np 6 2 device calib. Param.
65 129 F32+RW Sarasota2 Tcal 6 2 device calib. Param.
66 131 F32+RW Sarasota2 Pcal 6 2 device calib. Param.
67 133 F32+RW Input in UFP (if enabled): temperature body 6 1 °C
68 135 F32+RW Input in UFP (if enabled): temperature process 6 2 °C
69 137 F32+RW Input in UFP (if enabled): temperature proving. external flowmet 6 8 °C
70 139 F32+RW Input in UFP (if enabled): temperature densitometer 6 2 °C
71 141 F32+RW Input in UFP (if enabled): pressure process 6 2 bar