Instruction Manual

4
1 Important information for installation: PLEASE NOTE !
Location and position as required,
but electrode axis X – • – • – • – X
must be approximately horizontal in a
horizontal pipe run.
Y terminal box or converter housing
Measuring tube must be completetly filled at all times.
Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating
instructions for the signal converter.
Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34: below 2.2g for compact flowmeters
in the frequency range of 20-50 Hz with the IFC 010 K
/ IFC 020 K and 20-150 Hz
with the IFC 090 K.
Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.
Large meter sizes (
DN 200 /
8”): use adapter pipes to allow axial shifting of the
counterflanges and to facilitate installation.
X
X
Use only solventless detergents to clean the signal converter housing (polycarbonate).
Temperatures
Refer to Section 11 ”Limits” for operating pressure and vacuum load based on flange
standards and type of tube liner.
Ambient temperature Process temperature
Compact systems -25 to +60 °C (-13 to +140 °F) -25 to +60 °C (-13 to +140 °F)
-25 to +40 °C (-13 to +104 °F) -25 to > +60 °C (-13 to > +140 °F)
IFS 4000 F
IFS 4005 F -25 to +60 °C (-13 to +140 °F) -25 to > +60 °C (-13 to > +140 °F)
In storage -25 to +60 °C (-13 to +140 °F) with liners made of
Teflon
®
-PFA, Teflon
®
-PTFE, FEP, Tefzel, Irathane and soft rubber
-20 to +60 °C (-04 to +140 °F), kept immobile,
with Neoprene liner
Transport -25 to +60 °C (-13 to +140 °F), with liners made of
Teflon
®
-PFA, Teflon
®
-PTFE, FEP, Tefzel, Irathane and soft rubber
-05 to +50 °C (-04 to +140 °F),
with Neoprene liner
Teflon
®
is a registered trademark of Du Pont
Y