Owner manual

Part B IFC 040 Signal converter Section 4.4
02/2001
ALTOFLUX 2W
29
Display texts
Description and settings
FCt. 3.3
ZERO POINT
Calibrate the zero
Edit
not Sure
YeS Sure
Carry out only after a replacement of the electronic unit or if,
at low flow, an offset is presumed.
Please note!
Measuring tube must be completely filled with the
process liquid!
Flow must truly be “zero”!
8 XXX.X
Perc.
Display of current flow rate as a percentage of the full-scale
range. (The segments of the “8” are reduced in keeping with
the progress of measurement.)
Edit
not Save
Yes Save
Save new zero value?