Section 4.4 Part B IFC 040 Signal converter
ALTOFLUX 2W
02/2001
26
Display texts
Description and settings
FCt. 1.7
STATUS OUTPUT
Status output
Settings for the status output
Only possible when “Status output” set under Fct. 3.5
Hardware.
When “Pulse output” selected, Fct. 1.7 has “no function".
inFo
Function
Edit
off
on
all error
SiGn
overflow
emPtY PiPe
auto.ranGe
(für Stromaus
g
an
g
)
limit value
Standard: off
Setting the characteristic of the pulse output
Switch open Switch closed
permanently -
- permanently
error no error
positive flow negative flow
no overranging overranging
pipe/tube completely filled pipe/tube empty
range above limit range below limit
normal function zoom-in function active
inactive active
inFo
limit
XXX
Perc.
Range 5 ... 80%
Standard: 20%
When automatic range change selected
the limit value must be set as a percentage of the full-scale
range (Q
100%
), (see Fct. 1.1):
Below the set limit value, the current output has a zoom-in
function. The flow range from “0” to “Limit” is projected to the
I
0%
to I
100%
range.
For characteristic of automatic range change, see Fig. 4 in
Section 2.3
inFo
1.limit.value
XXX
Perc.
Range 0.1 ... 110 %
Standard: 10 %
inFo
2.limit.value
XXX
Perc.
Range 0.1 ... 110 %
Standard: 20 %
When limit value selected
the ‘on’ and ‘off’ values must be set as a percentage of the
full-scale range (Q
100%
) (see Fct. 1.1):
with adjustable hysteresis, as the ‘on’ value can be smaller or
larger than the ‘off’ value.
For characteristic of limit switch, see Fig. 5 in Section 2.3