Owner's manual

ACM 500 02/2013
10
Adjustable measuring ranges conductivity:
No. Range Resolution
1 0... 0.5 ms 0.001 ms
2 0... 1 ms 0.001 ms
3 0... 2 ms 0.010 ms
4 0... 3 ms 0.010 ms
5 0... 5 ms 0.010 ms
6 0... 10 ms 00.10 ms
7 0.. . 20 ms 00.10 ms
8 0... 30 ms 00.10 ms
9 0... 50 ms 00.10 ms
10 0...100 ms 000.1 ms
11 0...200 ms 000.1 ms
12 0...300 ms 000.1 ms
13 0...500 ms 000.1 ms
14 0...999 ms 000.1 ms
Ranges set in as-delivered condition:
Meas. range mS
1 0...200
2 0...20
3 0...2
4 0...0.5
3.4 Setting of temperature compensation 2, 4, 6, 8
Each measuring range is assigned its own setting for temperature compensation. The range is
adjustable between 0%/K (no compensation) and max. 5%/K. The compensation calculator
operates linearly on the basis of a reference temperature of 25°C. When supplied, the setting for
all ranges is 2%/K.
1
mS
Setting temperature coefficient
20
40
60
80
Meas. range No.
0... %/K
Measured conductivity
1
mS
Setting for TC 0...5.0 %/K
(with Jog Shuttle)
21
41
61
81
Meas. range No.
%/K
^
v
Measured conductivity