User Guide

70
Measurement
Preparation
(3) Copy from other calibration channel
7.
Press either
or
key to select
[CH COPY] and then
ENTER
key.
< COPY TO ??> screen appears.
8.
Press either
or
key to select
calibration channel to copy from.
To change calibration channel to copy to, press
SHIFT key to switch to SHIFT mode or press
or key to select calibration channel to copy to.
Once copied, calibration channel to
copy to cannot be returned to previous
status before coping. Make sure to check
channel number before copying.
9.
Press
ENTER
key.
After PLEASE WAIT...” appears and the
value is copied,<COPY TO> screen appears.
When PLEASE WAIT...” appears, do not
switch OFF. Doing so may break stored data.
continuation from a p.64, 65.
11 9 7, 8
USER CAL
<COPY TO>
CH02:<DEFAULT>
Lv ------
x ------
y ------
OK TO COPY ?
OK : [ENTER]
CANCEL : [ESC]
[ ] [ ]
<SOURCE>
CH
01
:
Lv 100.76
x 0.3945
y 0.3998
<
CH SETTING
>
CH02:
<
DEFAULT
>
USER CAL
CH ID NAME
CH COPY
CH RESET
BREAK:[ESC]
[ ] [ ]
<COPY TO>
CH
02
:<DEFAULT>
Lv ------
x ------
y ------
OK TO COPY ?
OK : [ENTER]
CANCEL : [ESC]
[ SFT] [ 2° ]
<SOURCE>
CH01 :
Lv 100.76
x 0.3945
y 0.3998
<COPY TO>
CH
02
:<DEFAULT>
Lv 100.76
x 0.3945
y 0.3998
OK TO COPY ?
OK : [ENTER]
CANCEL : [ESC]
[ SFT] [ 2° ]
<SOURCE>
CH01 :
Lv 100.76
x 0.3945
y 0.3998