Error message Cause (Description) Corrective Action
7
OVER
Luminance of measuring
object is higher than avail-
able measurement range.
Use ND filter and redo measure-
ment.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
8
OFFSET ERROR
Zero calibration has not
been correctly performed.
Reset the power and redo meas-
urement.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
9
EXCESSIVE LUMI-
NANCE VARIATION
Luminance change for
measuring object is large.
Place this instrument on stable
surface and redo measurement.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
10
VIEWING-ANGLE
SELECTOR ERROR
Measurement was per-
formed when measurement
angle selector was in wrong
position, or measurement
angle selector was changed
during measurement.
Switch measurement angle selec-
tor and check that measurement
angle appears in screen (------ is
not displayed), and then redo
measurement. In addition, do not
operate measurement angle selec-
tor during measurement.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
11
ROM ERROR
Data stored in ROM is
broken.
Do not switch OFF (O side) while
storing data or changing setting, or
when message “PLEASE WAIT...
appears.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.