User Guide

107
Description
Operation Procedure (Without data management software CS-S10w)
Setting Necessary for Object Color Measurement
1.
Set one or more tungsten lumps
or equivalent as illumination
source toward white calibration
plate as in the right illustration.
Set this instrument vertical
to white calibration plate.
Keep the angle between
illumination light source and
white calibration plate at 45˚.
White Calibration
2.
Perform user calibration.
See p.64 for details.
Measurement of Object
3.
Set object on the same position and at the
same angle as white calibration plate.
4.
Now, measurement can be done.
TARGET
COLOR
PEAK/VALLEY
SHUTTER
MEMORY
MEAS SPEED
ABS/DIFF
BACKLIGHT
KEY LOCK
CHAR MODE
RECALC
SHIFT
MENU ESC
ENTER
Illumination
Light Source
Illumination
Light Source
45°
45°
White calibration plate