Camera Accessories User Manual

112
Description
Error message Cause (Description) Corrective Action
12
MEMORY
ERROR
Data stored in ROM is
broken, or backup batteries
drain.
Do not switch OFF (O side) while
storing data or changing setting, or
when message “PLEASE WAIT
appears.
Switch ON (| side) to charge back-
up battery. Backup battery is to
fulfill its function after fully charged
for approx. 20 hours.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
13
INCORRECT
CLOCK
OPERATION
Clock IC does not operate
correctly.
Reset the power to set correct
date and time.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
14
A/D ERROR Error in AD converter Reset the power to redo measure-
ment.
If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.
15
RECALCULA-
TION ERROR
There exists no measure-
ment value for recalculation,
or measurement value be-
comes lower than measure-
ment available range after
recalculation.
Check that measurement value ap-
pears and redo measurement.
Check that user calibration is cor-