7-6 A-61628 December 2011 
6. Reinstall the feed module by aligning the pins and pushing it toward 
the left to fit it into position. Be sure the gears are aligned and it 
snaps into place. Roll the feed module back into place.
7. Reinstall the feed module cover by aligning the tabs with the slots 
and pushing it back into position until it snaps into place.
Vacuuming the Scan 
Station
Before cleaning the imaging area, it is suggested that you vacuum the 
inside of the Scan Station to remove any dust and debris.
Cleaning the imaging area 1. Wipe the upper and lower imaging area using a Staticide wipe. Be 
careful not to scratch the imaging area when cleaning.
IMPORTANT: Staticide wipes contain isopropanol which can cause eye 
irritation and dry skin. Wash your hands with soap and 
water after performing maintenance procedures. Refer to 
the MSDS for more information.
2. Wipe the upper and lower imaging area again using the reusable 
optics cleaner pad to remove any streaks. 
3. When finished, close the Scan Station cover with both hands.