Zero Adjust
A zero adjustment is normally not needed. In case, that the unit does not indicate U=00X while the
modulator system is shielded (e.g. grounded person covers the modulator system with one hand),
use the trimmer (6) for zero adjustment. The last digit can be ignored as the caused measurement
failure is much lower than the specified tolerance.
Guaranty Performance
By correct operation and in accordance to the instruction manual we allow a 24 month warranty .
The warranty does not include: The Battery / NiMH – accu, damages by electrical discharge,
incorrect grounding and mechanical damage of the Field Meter.
If you opened the unit, you will lose the Guaranty !
Warning notice
√ It is not allowed to open the Electrofieldmeter. If you open it, you lost the guaranty.
√ The Electrofieldmeter is not approved for measurements in explosion hazard areas.
√ In the event of very high electrostatic charges, the Electrofieldmeter must be grounded. Look for
an adequate distance to the test object.
√ Avoid spark discharges onto the modulator system.
√ The use of the unit in power plants is not permitted.
√ The Electro-Fieldmeter is not equipped to detect alternating fields > 1Hz.
Calibration
It is advisable to recalibrate the Electro-Fieldmeter once a year.