6
With the Charge Plate Kit becomes from the EFM 022 a charge plate monitor , to the supervision
of air ionization appliances after the EN 100015 as well as EOS/ESD S.3.1
The charge electrode corresponds to the norms from the capacity (20pF), but not from the
dimensions. For directly measure conform to the norms, use our CPM 374.
With the CPS 022, however reproducible measurements can be performed.
It becomes the decay-time from + 1.000V…. + 100V or.- 1.000V....- 100V with an integrated
start - stop - timer. The charge takes place over an external HV- Converter.
In the CPS-Mode voltages up to max. 2kV will be displayed.
Operating instruction CPS022
Parts of delivery
• HV Converter
• 3 rechargeable batteries
• Battery charger
• CPS Socket
• Plate condenser
• Grounding cables (1.5m & 1m)
• Crocodile plug
• Suitcase