Instructions / Assembly

139704 Rev. 09/14 B
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
customerservice@kleintools.com
www.kleintools.com
ENGLISH
ENGLISH
• NON CONTACT
VOLTAGE TESTER
• TRUE RMS
• AUTO / MANUAL
RANGE
• DATA HOLD
• MAX / MIN
• BAR GRAPH
• BACKLIGHT
• WORKLIGHT
• 3-5/6 DIGIT
5999 COUNT LCD
2200
SYMBOLS USED ON LCD
AC Measurement DC Measurement
Negative DC Value
AT
Auto Range Active
O
.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
M
AX
Maximum Reading
%
Duty Cycle Mode Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms Diode Test
F
Capacitance in Farads Continuity Test
Relative / Zero Mode
n
Nano 10
-9
m
Milli 10
-3
µ
Micro 10
-6
M
Mega 10
6
k
Kilo 10
3
8. Non Contact Voltage (NCV): > 25V AC
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range Resolution Accuracy
600mV to
600V
0.1mV to
100mV
± (0.5% + 4 digits)
1000V 1V ± (0.8% + 10 digits)
Overload Protection: 1000V
AC Voltage Measurement
Range Resolution Accuracy
600mV to
750V 0.1mV to
1V ± (2.0% + 5 digits)
Overload Protection: 750V RMS
Frequency Response: 45 to 400 Hz
Minimum Voltage for Frequency Measurement: 200mV
Response: True RMS
DC Current Measurement
Range Resolution Accuracy
600µA 0.1µA
± (1.2% + 3 digits)
2000µA 1µA
60A 0.01A ± (2.5% + 15 digits)
600A 0.1A ± (1.5% + 8 digits)
Overload Protection:
Voltage: 600V RMS
Current: Leads 2000µA, Clamp 600A
AC Current Measurement
Range Resolution Accuracy
600µA 0.1µA ± (2.0% + 5 digits)
2000µA 1µA ± (1.5% + 5 digits)
60A 0.01A ± (2.9% + 15 digits)
600A 0.1A ± (1.9% + 8 digits)
Overload Protection:
Voltage: 600V RMS
Current: Leads 2000µA, Clamp 600A
Frequency: 45 to 400Hz
Minimum Current for Frequency Measurement: 400µA or 20A
Minimum Current for Clamp Measurement: 0.2A
Response: True RMS
Resistance Measurement
Range Resolution Accuracy
600
Ω to 6MΩ 0.1Ω to 1kΩ
± (1.0% + 4 digits)
60M
Ω 10kΩ
± (2.0% + 4 digits)
Overload Protection: 600V RMS
Capacitance Measurement
Range Resolution Accuracy
60nF to 6000µF 0.01nF to 1µF ± (3.5% + 6 digits)
Overload Protection: 600V RMS
Frequency Measurement (Voltage)
Range Resolution Accuracy
99.99Hz to 499.9kHz 0.01Hz to 100Hz ± (0.1% + 4 digits)
Overload Protection: 600V RMS
Sensitivity: 2.5V RMS
Duty Cycle Measurement
Range Resolution Accuracy
1 to 99% 0.1% ± (0.1% + 5 digits)
Overload Protection: 600V RMS
Sensitivity: 1.2Vpp ~ 50Vpp
Frequency Range: < 500 Hz
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS
6.0V Appx. 0.25mA
<
1.8V DC
Continuity Test
Overload Protection Open Circuit Voltage
600V RMS
<
0.44V
Non Contact Voltage Detector
On Voltage
Appx. 25V AC
WARRANTY
www.kleintools.com/warranty
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
750V
1000V
600A
PRESS &
HOLD
PRESS &
HOLD
Dwg Name: CL2200-139704ART Dwg No: 139704
ECO No: 17821 Pkg Dwg Ref: 1650 Rev: B
Color Reference: N/A
CL2200
INSTRUCTION MANUAL
600A AC/DC TRMS Clamp Meter
CL2200 Instructions - MASTER.indd 1-4 9/25/2014 11:48:21 AM

Summary of content (8 pages)