User Manual

Resistance measurement
range accuracy
(1)(3)
Resolution /
noise
(1)
linearity
(2)
temperat.
drift
(2)
20 k 10
10 k 5
5 k 2.5
0.003 %
2.5 k 1.25
0.05 %
1.25 k 0.63
0.01 % 25 ppm/K
625 0.31
312.5 0.15
0.010 %
5 ppm/K
(4)
200 0.1
0.003 %
(1) values for T=20 °C , (2) values estimated from the
hardware specification
(3) values correspond to 80 % of the measuring range
(4) with external drift compensation, reference resistance
with TCR 1
Achievable absolute accuracy with voltage measuring incl. drift correction:
typical ±3 µV
Achievable accuracy for temperature measurement with Pt 100 resistance
(measuring range -200°...850 °C, temperature range -200°...200 °C, scan
rate 1 s, averaging interval 1 min)
without drift correction: 0.1 °C with ambient temperatures
0 °C...+40 °C
with external drift correction: typical 0.05°C with ambient temperat-
ures
-40 °C...+85 °C
Reference resistance 100 , TCR 1
Hardware Manual COMBILOG 1022
167