Accuracy: 0.05 %, (dependent on range)
Resolution: 0.003 .. 0.03 %,
(dependent on range)
Linearity: 0.01 %
Temperature drift: 25 ppm/K
5 ppm/K with additional (ext.) drift
correction
13.5 Digital Inputs/Outputs (6 per Module)
as input:
Function: status, frequency, counter,
8-bit Graycode receiver
Input voltage: max. +30 VDC
Input current: max. 1.5 mA
Input frequency: max. 4000 Hz
Switching threshold (low): > 3.5 VDC
Switching threshold (high): < 1.0 VDC
Counter width: 16 bit
as output:
Function: host-out, process-out
Output voltage: max. 30 VDC
Output current: max. 100 mA
Type of output: open collector
13.6 Interfaces
Base: RS 485, RS 232, Ethernet, USB
RS232 and RS485 galv. isolated
from processor core, no isolation
between channels
Data format: 8E1 / 8N1 / 8O1
Protocol: ASCII, PROFIBUS, MODBUS
Hardware Manual COMBILOG 1022