User Manual
device identification (v...v) length = 28 char
<vendor name> ASCII ("Friedrichs") 10 char
<model name> ASCII ("COMBILOG") 8 char
<hw-revision> ASCII ("xy.yy") 5 char
<sw-revision> ASCII ("xy.yy") 5 char
x ... “M“ : monitor program
x … ”A”,“U“: universal program
x ... “T“ : calibration and test program
x ... “S“ : application specific program
y.yy : version
device information (s...s) length = 28 char
<location> ASCII 20 char
<serial number> ASCII 6 char
<number of channels> ASCII 2 char
status information (z...z) length = 12 char
< channel status > ASCII 8 Char
< module status > ASCII 4 Char
<channel status> = K32..K29 K28..K25 … K8..K5 K4..K1
Byte: 1 2 … 7 8
<module status> = M16..M13 M12..M9 M8..M5 M4..M1
Byte: 9 10 11 12
If the bit Kn in the channel status is set it indicates that an error
has occurred in channel n. A channel error is given when the
measuring value is outside of the linearisation, e.g. in case of a
sensor break down or of a short circuit of transmission.
Hardware Manual COMBILOG 1022
113