By Paul Aylett at 3:43 pm, Jun 14, 2012 Non-contact Forehead Thermometer User Manual NCT1 Series
CONTENT Introduction 2 Cleaning and Storage 12 Parts 3 Specification 13 Safety Precautions 4-6 Problems & Troubleshooting 14 Setting Up Your Non-contact Forehead Thermometer 7 Warranty 15 Using Your Non-contact Forehead Thermometer 8-11 Explanation of Symbols on Unit 16 Measurement Process 8 EMC Table 17-22 Functions 9-10 Replacing Battery 11
Introduction Thank you for purchasing The Kinetik Non-contact Forehead Thermometer (Model NCT1). For optimal use of this product, please read the User Manual before any measurements are taken. For safe and proper use of this product, please read and fully understand the Safey Precautions in this User Manual. Keep this User Manual for easy reference.
Parts Product Structure: Sensor Lens Battery Last reading LCD Battery Cover Release Hole ON/MEMORY Room temperature Surface temperature Battery Cover Scan Forehead temperature Memory 3
Safety Precautions Warning marks and symbols are indicated for safe and proper use of this product and prevention of any injury to you and others. See Table below for description of warning marks and symbols: Read operating instruction. WARNING Means a possibility of personal injury if the thermometer is not used correctly. NOTICE Means a possibility of personal injury or property damage* if the thermometer is not used correctly.
Safety Precautions WARNING MUST OBSERVE It is not advisable for users to make a self – diagnosis and self treatment based on the temperature results obtained. Seek medical advice. A self – diagnosis may result in deterioration of the condition. Do not touch or blow the infrared sensor. A dirty infrared sensor may lead to reading inaccuracy. Ensure the infrared sensor is cleaned before use for accurate reading. Clean the infrared sensor carefully with a cotton swab or soft cloth moistened with alcohol.
Safety Precautions NOTICE PROHIBITED Do not use the Non-contact forehead thermometer for purposes other than those for which it was designed. Do not drop, stand on or subject the Non-contact forehead thermometer to vibration or impact. ATTENTION Dispose battery with care. Do not throw battery in fire.
Setting Up Your Non-contact Forehead Thermometer Preparations: 1. To activate the Non-contact forehead thermometer, pull out the insulated strip. 2. For accurate reading, condition the thermometer at room temperature for 30 minutes before use. 3. An unexpected change in ambient temperature is likely to reduce the reading accuracy. Any attempt to perform temperature measurment in front of an air conditioning unit will result in inaccurate readings. 4.
Using Your Non-contact Forehead Thermometer Measurement Process: 1. Always make sure the probe lens is clean without any damage and the forehead is clean. 2. Power on: Press the ON/MEMORY button (see figure 1). 3. Measuring body temperature on the forehead: Forehead mode is the default mode. You can see the icon on the screen and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer 1 to 3 cm from the central forehead (Fig. 2) and press the "Scan" button.
Using Your Non-contact Forehead Thermometer Functions: Forehead temperature The thermometer has been designed for practical use. It’s not meant to replace a visit to the doctor. Please also remember to compare the measurement result to your regular body temperature. Please consult with doctor if you have health concerns. Please see the “Using Your Non-contact Forehead Thermometer” section to learn how to measure the body temperature.
Using Your Non-contact Forehead Thermometer Functions: Last reading After getting a new temperature reading, the last reading will be shown on the screen (at top right corner) with the icon. Memory locations There are 25 memories for forehead measurements. After power on, press the ON/MEMORY button to see the temperature records with the icon. °C /°F switch In “Power Off” mode, press and hold the “SCAN” button, then press the ON/MEMORY button for 3 seconds, icon “°C” will be switched to icon “°F”.
Using Your Non-contact Forehead Thermometer Replacing Battery : This device is supplied with one lithium cell CR2032x1. 1. Open the battery cover: Gently insert a small pointed object such as a watch-makers screwdriver into battery cover release hole. At the same time, use thumb to push battery cover out. 2. Hold the device and flip the battery out with a small screwdriver. 3. Insert a new battery under the metal hook and press down until you hear a “click” sound. 4.
Cleaning and Storage: Please make sure the probe is clean to ensure an accurate reading. The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage. a. Use the alcohol swab or the soft cotton moistened with the alcohol to clean the probe lens. b. Allow the probe to fully dry for at least 1 minute. c. Keep the unit dry and away from any liquids and direct sunlight. d.
Specification Temperature measurement range: Forehead mode: 34~42.2°C (93.2~108°F) Surface mode: -22~80°C (-7.6~176°F) Operating temperature range: 10~40°C (50~104°F) Storage temperature range: It should be stored at room temperature between -20~+50°C , RH≦85% Transportation temperature shall be less than 70°C, RH≦95% Comply with ASTM E1965-98, IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards. Accuracy: Forehead mode: ±0.2°C for 36.0~39.0°C, others ±0.3°C (0.
Problems & Troubleshooting Error Message Problem Solution Error 5~9, the system is not functioning properly. Unload the battery, wait for 1 minute and repower it. If the message reappears, contact the retailer for service. Measurement before device stabilization. Wait until all the icons stop flashing. The device showing a rapid ambient temperature change. The ambient temperature is not within the range between 10°C and 40°C (50°F ~104°F). (1) In forehead mode: Temperature taken is higher than +42.
Warranty Accessories: 1 x User Manual Warranty: One year warranty is available from purchasing date, excluding user-caused failures listed below: 1. Failure resulting from unauthorised disassembly and modification. 2. Failure resulting from unexpected dropping during application or transportation. 3. Failure resulting from not following the instructions in the user manual.
Explanation of Symbols on Unit 0197 Caution Please read the instructions for use. Manufacturer BF type applied part Battery Recycling Paper Recycling Harvard Medical Devices Ltd. Unit 1301-2, 13th Floor, Railway Plaza, 39 Chatham Road South, Tsimshatsui, Kowloon, Hong Kong. Kinetik Medical Devices Ltd.
EMC Table Guidance and manufacturer’s declaration – electromagnetic emissions The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the NCT1 should assure that it is used in such an environment.
EMC Table Guidance and manufacturer’s declaration – electromagnetic immunity The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the NCT1 should assure that it is used in such an environment.
EMC Table Guidance and manufacturer’s declaration – electromagnetic immunity NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a.
EMC Table Guidance and manufacturer’s declaration – electromagnetic immunity The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the NCT1 should assure that it is used in such an environment.
EMC Table Guidance and manufacturer’s declaration – electromagnetic immunity Immunity test interruptions and voltage variations on power supply input lines IEC 60601 Compliance level Electromagnetic environment – guidance test level <5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) IEC 61000-4-11 for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m Not applicable 3 A/m Mains power quality s
EMC Table Recommended separation distances between portable and mobile RF communications equipment and the ME EQUIPMENT or ME SYSTEM The NCT1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.